该产品已停产。可能仍有剩余库存。请参阅下面 "类似产品 "下的替代产品或直接联系我们
此停产产品的替代产品:
X1-4
Broadband radiometer for radiation protection measurements of UV radiation devices
- Mobile ligh meter for UV radiation while working with artificial UV (UVSV)
- Erythema effective irradiance and detection of UV-C radiation in accordance with the UV protection Act for solaria (sunbeds).
- UVSV, DIN EN 60335-2-27 and DIN 5050-1:2010-01 conformity
European Commission UV Protection Directive
UV Erythema Broadband Radiometer
UVA and UVB Two-channel UV Erythema detector head
Actinitic UV Erythema Dose-Response Function
Detection of UVC Radiation
UV Protection Directive Conformity
Artificial UV Sources and Sun Spectrum Correction
Cosine-Corrected Field of View
Simple Operation and Handling
Usage with a Computer
Threshold values from the UV Protection Directive
Relevant norms and regulations: UV Protection Directive, DIN EN 60335-2-27 (VDE 0700-27):2009-04, DIN 5050-1:2010-01
Threshold values irradiance | |
UV Protection Directive | 250 nm - 400 nm 0.3 W/m²eff 200 nm - 280 nm 0.003 W/m² |
60335-2-27 | 250 nm - 320 nm 0.15 W/m²eff 320 nm - 400 nm 0.15 W/m²eff |
5050-1 | No threshold value defined for UVC 200 nm - 290 nm |
Threshold values spectral correction | |
UV Protection Directive | Consult 60335-2-27 and 5050-1 |
60335-2-27 | No value for spectral correction |
5050-1 | f1< 0.2 for broadband radiometer without specification of reference spectrum |
Threshold value cosine correction | |
UV Protection Directive | Consult 60335-2-27and 5050-1 |
60335-2-27 | No specifications for difference to cosine |
5050-1 | f2 <0 .06 for broadband radiometer |
Threshold value temperature coefficient | |
UV Protection Directive | Consult 60335-2-27and 5050-1 |
60335-2-27 | Measurement at 20 °C ± 5 °C ambient temperature |
5050-1 | 0.005 K-1 (10 °C - 50 °C) |
Burn-in period of UV Emitters | |
UV Protection Directive | Consult 60335-2-27and 5050-1 |
60335-2-27 | Half of the maximum exposure time |
5050-1 | At least 15min and fluctuations less than ± 5 % |
Exposure time for 100J/m² | |
UV Protection Directive calculation example | At Eeff 0.3 W/m² about 5 min und 30 s 100 J/m² / 0.3 W/m² ≈ 330 s |
Threshold value measurement distance | |
UV Protection Directive | Consult 60335-2-27and 5050-1 |
60335-2-27 | Min. recommended distance 100±2mm for facial radiation devices |
5050-1 | Full body radiation 0mm or in a radius of 30 cm in front of the measurement point on the support surface |
X14 with BHO case
Display of the UV Erythema effective and UVC irradiance with selected correction factor for solar radiation at 70 ° zenith angle
X14 being used for the measurement of UV radiation due to solar radiation.
XD-45-ERYC-4 UV erythema detector head with additional UVC radiation detection function